Testing Applications
- Nuclear air and Gas Treatment Systems (Safety and Non-Safety Grade)
- High Efficiency Carbon (Charcoal) Air Filtration Systems
- Lab Fume Hoods/Exhaust
- Vapor Removal/ Scavenger Systems
- CFC Detection
The NUCON Model SAE is the only continuous Halide detector available that meets ANSI/ASME standards N510 and AG-1. The model SAE can be calibrated for Vppb level detection of two most commonly used challenge Halides (R-11 and R-12).
- Switchable Vppm/Vppb Ranges
- Direct Continuous Real Time Readout
- External "Zero" AirFilters For Easy Change-Out
- Automatic Saple Flow Control
- Recorder Output
- Solid-State Electronics
*R-11, R-12, SF6 and others
Technical Specifications and Information:
Sensitivity to Halides:
1 Vppb
Sampling Volume:
2000 ml/min
(Electronically Stabilized)
Linear Range:
1Vppb to 40 Vppm
Response Range:
1 Vppb to 80 Vppm
Accuracy:
±10%
Recorder Output:
2.0 V. 1mA
(Short-circuit protected)
Size:
5.3" (13 cm) H x 18" (44 cm) L
x 10.6" (27 cm) D
Weight:
23 lbs. (10.5 Kg)
Power Requirements:
100VA, 120VAC ±10%, 60Hz.
(220VAC, 50 Hz also available)